5 properties - calibration, Properties - calibration, Calibration – Metrohm viva 1.1 (ProLab) User Manual
Page 897
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
viva 1.1 (for process analysis)
■■■■■■■■
885
Number of pixels
Shows the resolution with which the spectrum is recorded.
7.6.5
Properties - Calibration
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Calibration
The Calibration tab shows the data of the wavelength calibration. A
wavelength defined by four wavelength coefficients is allocated to each
pixel on the detector. This wavelength calibration is carried out with the
CAL Spec command. In this process, the calibration data is calculated
automatically and displayed on this tab.
Date
Shows the time and date of the wavelength calibration.
User
Shows the user logged in at the time of the wavelength calibration.
Method
Shows the method with which the wavelength calibration has been car-
ried out.
Selection
Factory settings | "Method name"
Default value
Factory settings
[Reset calibration]
Resets the device to the factory-set calibration. In this process, Dark
spectrum and Reference spectrum are deleted.
Determination ID
ID of the determination with which the wavelength calibration has been
carried out. With the Factory settings calibration method, ---- is dis-
played.
The determination ID can be copied and pasted into another field.
Calibration coefficients
Coefficients of the regression function with which the CCD pixels are allo-
cated to a wavelength.
c#
Shows the wavelength coefficients c1 - c4.