4 add sample, 1 add sample - overview, Add sample – Metrohm viva 1.1 (ProLab) User Manual
Page 467
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
455
CAUTION
The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillaries with a diameter < 1 mm.
Filling rate
Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the dosing unit used and on the diameter of the tubing
at the selected fill port.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
5.5.2.4.4
ADD SAMPLE
5.5.2.4.4.1
ADD SAMPLE - Overview
Dialog window: Method
▶ ADD AUX ▶ Properties... ▶ ADD SAMPLE - 'Command
name'
Command for the addition of the sample.
Appearance
The command has the following appearance:
Parameters
The parameters for the ADD SAMPLE command are configured in the
following dialog window:
■
Command variables
The following command variables are generated by the ADD SAMPLE
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started