2 swing - properties – Metrohm viva 1.1 (ProLab) User Manual
Page 443
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
431
Command variables
The following command variables are generated by the SWING command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
SAN
Current absolute swing angle of the robotic arm in °
(entry upon exiting the command)
SPO
Current external position (entry when ending the
command); 0 means invalid position
5.5.2.3.3.2
SWING - Properties
Dialog window: Method
▶ SWING ▶ Properties... ▶ SWING - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Instrument type field can no longer be
edited and the instrument type belonging to the instrument is displayed.