5 editing baselines – Metrohm viva 1.1 (ProLab) User Manual
Page 673
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
661
Modifies the sequence of shown substances by moving the selected line
downward.
Edit menu
Properties...
Edit baseline parameters of the selected substance (same action as double-click
on the line). The properties window opens for entering the baseline parameters
(see Chapter 5.6.3.5, page 661).
5.6.3.5
Editing baselines
Dialog window: Method
▶ Evaluation ▶ Substances ▶ Base line ▶ [Edit] ▶ Proper-
ties...
▶ Substances - Baseline
Measuring command
Shows the measuring command for the substances of which the baselines
can be defined.
Substance
Shows the substance name.
Base line
Baseline type
Selection of the baseline type with which the evaluation is to take place.
Selection
horizontal | Horizontally through 0
Default value
horizontal
horizontal
Forced, horizontal baseline starting from the automatically determined
or manually specified end base point.
Horizontally through 0
Forced, horizontal baseline through point of origin.
Base point evaluation
Selection of the method for determining the base points.
CVS
Selection
Automatically | Manual
Default value
Automatically
Automatically
If this option is selected, then the base points are determined automat-