2 curve properties va measuring command - y1 axis – Metrohm viva 1.1 (ProLab) User Manual
Page 187
■■■■■■■■■■■■■■■■■■■■■■
3 Workplace
viva 1.1 (for process analysis)
■■■■■■■■
175
decimal
The axis is labeled in the decimal format ±##.### (floating-point num-
ber).
scientific
The axis is labeled in the scientific format ±#.###e±###.
engineering
The axis is labeled in the technical format ±###.###+prefix (engi-
neering format). The prefix (m, µ, n, p) is added to the figure.
Scaling
on | off (Default value: off)
If this check box is activated, then the x axis can be manually scaled in the
curve window. In this case, the Start value and End value fields can be
edited. If this check box is not activated, then the x axis will be automati-
cally scaled.
Start value
Initial value for scaling the x axis.
Editable only when autoscaling is disabled.
Input range
-1.00E+12 - +1.00E+12
Default value
-1.00E+12
End value
End value for scaling the x axis.
Editable only when autoscaling is disabled.
Input range
-1.00E+12 - +1.00E+12
Default value
+1.00E+12
3.9.2.2.2
Curve properties VA measuring command - y1 Axis
Tab: Workplace / Database
▶ Curves # ▶ Properties Curves # ▶ y1 Axis
Parameters for the graphical display of the curves on the y1 axis (left-hand
y axis).
Quantity
Selection of the quantity to be shown on the y1 axis.
CVS
Selection
Current [A] | Potential [V] | Time [s] | Current
measuring range [A]
Default value
Current [A]