2 add sample - properties – Metrohm viva 1.1 (ProLab) User Manual
Page 468
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
456
■■■■■■■■
viva 1.1 (for process analysis)
Identification
Description
CYL
Cylinder volume of the dosing unit used for the com-
mand
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
VOL
Currently dosed volume in L (Number)
5.5.2.4.4.2
ADD SAMPLE - Properties
Dialog window: Method
▶ ADD SAMPLE ▶ Properties... ▶ ADD SAMPLE - 'Com-
mand name'
Command name
Name of the command.
Entry
25 characters
Addition
Selection of the addition mode for the sample.
Selection
Add manually | Already added | Add with dos-
ing device
Default value
Add manually
Add manually
Manual addition of the sample, at which time the track is stopped. A
message appears and an acoustic signal is heard.
Already added
The sample is located already in the measuring vessel or was channeled
into the measuring vessel, e.g. with a peristaltic pump. The track is not
stopped. No message appears and no acoustic signal is heard.
Add with dosing device
Automatic addition of the solution with a dosing device. The Sample
amount is added for undiluted samples, the Analytical volume for
diluted ones. The sample is added through the Dosing port 1 defined
in the configuration for the dosing unit (see Chapter 6.6.3.2, page
785).