Metrohm viva 1.1 (ProLab) User Manual
Page 681
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
669
Calibration method RC
Substance
Name of the substance. Substances can neither be added nor deleted
here, this must take place in the substance table (see Chapter 5.6.3, page
655).
Evaluation quantity
Measured quantity for the evaluation of the calibration curve.
Curve type
Curve type for the evaluation of the calibration curve.
Evaluation start
Evaluation ratio Q/Q(0) with which the calculation of the regression for
the standards is to be started.
Weighting
Information as to how the points of the calibration curve are to be
weighted.
Calibration method LAT
Substance
Name of the substance. Substances can neither be added nor deleted
here, this must take place in the substance table (see Chapter 5.6.3, page
655).
Evaluation quantity
Measured quantity for the evaluation of the calibration curve.
Curve type
Curve type for the evaluation of the calibration curve.
Weighting
Information as to how the points of the calibration curve are to be
weighted.
Edit menu
Properties...
Opens the Properties calibration dialog window in order to edit the data of
the highlighted line.
Apply settings for all
substances
The settings of the highlighted line are adopted for all substances.