5 Method
viva 1.1 (for process analysis)
■■■■■■■■
643
Command variables
The following command variables are generated by the command CP in
the method run and can be used in formulas under the designation 'Com-
mand name.Variable identification':
Identification
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has
never been started
DBL
Total duration for the processing of the command in
s
ELT
Variable for electrode test:
0 = Electrode test has not been performed
1 = Electrode test has been performed once
2 = Electrode test has been performed twice
3 = Electrode test has been performed three times
FIN
Command status:
1 = Command has been called and ended at
least once
0 = Command has been called and never
ended
invalid (variable not available) = Command has
never been called
HR
Highest potential measuring range used:
1 = 5 V | 3 = 500 mV | 5 = 50 mV | 7 = 5 mV
LR
Lowest potential measuring range used:
7 = 5 mV | 5 = 50 mV | 3 = 500 mV | 1 = 5 V
NMP
Number of measuring points in measuring point list
RPM
Stirring rate in min
–1
STY
Type of stop with which the command was stopped:
1 = finished normally, 0 = stopped manually
with [Quit] or [Stop]