3 recalculation rules – Metrohm viva 1.1 (ProLab) User Manual
Page 297
■■■■■■■■■■■■■■■■■■■■■■
4 Database
viva 1.1 (for process analysis)
■■■■■■■■
285
This button is disabled for as long as recalculation has not yet been trig-
gered and if not all the selected determinations were able to be recalcu-
lated.
[Cancel] or
Closes the Reprocess dialog window. The result of the reprocessing
made since the last saving will not be saved.
Reprocessing with history
Versioning behavior of reprocessed determinations.
Selection
On | Off
Default value
Off
On
A new version is created and saved each time a determination is repro-
cessed.
Off
The version number is increased with each reprocessing of a determi-
nation, but only two versions are stored in the database (the first, origi-
nal version and the latest, reprocessed version). In the case of already
reprocessed determinations, the version that was last saved is deleted
and a new version is created.
4.5.2.5.3
Recalculation rules
Dialog window: Reprocess
The following rules apply for recalculating the selected and modified
determinations:
■
Number of determinations
No more than 10 determinations may be selected for recalculation.
■
Start test
During the start test all device tests and device monitoring will be
ignored.
■
Special tracks
Series start track, series end track and error tracks are not run through
during recalculation.
■
Tracks with "Return immediately"
If the method contains tracks with the option Return immediately
enabled, then a warning will will appear, because it can no longer be
guaranteed that during recalculation the method will run in the same
way as during the determination.