4 sample changer - assigning a position, Assigning position – Metrohm viva 1.1 (ProLab) User Manual
Page 983
■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
viva 1.1 (for process analysis)
■■■■■■■■
971
Input range
10 - 55 °/s
Default value
55 °/s
[Start]
Starts moving to the target position. After the start, the button changes to
[Stop], the two lower buttons are shown as inactive (gray) and Move... is
shown as status message instead of Ready.
Moves the robotic arm to the left (towards 0°) at a swing rate of 55°/s as
long as this button is pressed.
Moves the robotic arm to the right (towards 330°) at a swing rate of
55°/s, as long as this button is pressed.
8.4.5.4
Sample changer - Assigning a position
Program part: Manual control
Here you can assign the current rack position, lift position or robotic arm
position to a specific special position.
Rack position
Assign the current rack position to a specific special beaker.
Current position
Shows the current rack position.
Special beaker
Selection of the special beaker to which the current rack position is to be
assigned.
Input range
1 - 16
Default value
1
[Assign]
Triggers assignment. During the assignment the cursor is displayed as
hourglass.
Lift position
Assign the current lift position to a specific special position.
Current position
Shows the current lift position in mm.