Metrohm viva 1.1 (ProLab) User Manual
Page 497
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
485
Command variables
The following command variables are generated by the CAL MEAS Opt
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':
Identification
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has
never been started
DBL
Total duration for the processing of the command in
s
EME
End measured value (measured value after process-
ing of the command) in the unit of the measured
value
ETE
1)
End temperature (temperature after the command
has been processed) in °C
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
invalid (variable not available) = Command has
never been started
IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
ITE
1)
Initial temperature (temperature before start condi-
tions are processed) in °C
LP.CNT
Transmission of the last measuring point in the mea-
suring point list in %
LP.MEA
Measured value for the last measuring point in the
measuring point list in the unit of the measured
value
LP.TEM
Temperature for the last measuring point in the mea-
suring point list in °C
LP.TIM
Time in s until the last measuring point in the mea-
suring point list is reached