5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
502
■■■■■■■■
viva 1.1 (for process analysis)
Parameters
The parameters for the RECEIVE command are configured in the follow-
ing dialog window:
■
Command variables
The following command variables are generated by the RECEIVE com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
TOU
Timeout status: 1 = Max. waiting time expired; 0
= Max. waiting time not expired
5.5.2.6.5.2
RECEIVE - Properties
Dialog window: Method
▶ RECEIVE ▶ Properties... ▶ RECEIVE - 'Command name'
Command name
Name of the command.
Entry
25 characters
Wait for event/status
Selection
Wait until all conditions are met | Wait until
one of the conditions is met
Default value
Wait until all conditions are met
Wait until all conditions are met
A wait ensues until all of the conditions listed in the table are fulfilled.
Wait until one of the conditions is met
A wait ensues until one of the conditions listed in the table is fulfilled.
[New]
Opens the Wait for event/status dialog window for entering a new
event message which is to be waited for (see Chapter 5.5.2.6.5.3, page
503).