4 analog out - analog output, 5 analog in, 1 analog in - overview – Metrohm viva 1.1 (ProLab) User Manual
Page 575: Analog in
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
563
5.5.2.8.4.4
ANALOG OUT - Analog output #
Dialog window: Method
▶ ANALOG OUT ▶ Properties... ▶ ANALOG OUT - 'Com-
mand name'
▶ Analog outputs ▶ [Properties] ▶ ANALOG OUT - Analog output #
Port description
Selection of the analog output. The analog outputs of the selected instru-
ment are available. No port can be selected if not defined is selected in
the Device name field.
Type
Shows the analog output type.
Action
Selection
Set value | Set default
Default value
Set value
Set value
Writes the content of the Value field on the analog output.
Set default
Writes the content of the Default value field from the configuration
on the analog output.
Value
Mapped value that is to be written on the analog output. A fixed value or
a formula (formula editor, CTRL F or right mouse button) can be entered.
Mapping area and unit of the value are defined in the configuration of the
analog output.
If the value that is to be written is smaller than the 0% Mapping, then
the smallest possible value (Low level (0 %)) is output; in the case of val-
ues greater than the 100% Mapping, the greatest possible value (High
level (100 %)) is output.
5.5.2.8.5
ANALOG IN
5.5.2.8.5.1
ANALOG IN - Overview
Dialog window: Method
▶ ANALOG IN ▶ Properties... ▶ ANALOG IN - 'Command
name'
Instruments
This command can be executed with the IO controller.
Appearance
The command has the following appearance: