4 meas opt - evaluations, Evaluations – Metrohm viva 1.1 (ProLab) User Manual
Page 537
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
525
089 Photometer
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 4,000 mAU
Selection
off
Default value
off
off
No stop.
Temperature
This section is displayed only when Device type = 089 Photometer.
Temperature
Measuring temperature, which can be entered manually. The temperature
is measured continuously when Temperature measurement = contin-
uous (tab General/Hardware).
Input range
20.0 - 60 °C
Default value
25.0 °C
5.5.2.7.4.4
MEAS Opt - Evaluations
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ MEAS Opt - 'Command name' ▶ Eval-
uations
Command name
Name of the command.
Entry
25 characters
The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.
Fixed point evaluation
Fixed point evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value or Time) for the fixed point from the measuring point
list.