4 properties - glp, Properties - glp – Metrohm viva 1.1 (ProLab) User Manual
Page 911
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
viva 1.1 (for process analysis)
■■■■■■■■
899
Input range
0 - 9,999 ms
Default value
0 ms
[Connect]
Establishes the connection to the RS-232 device. The Establish connec-
tion dialog window appears (see Chapter 7.8.5, page 901).
[Disconnect]
Disconnects the connection to the RS-232 device (only then can the
device be deleted from the configuration).
7.8.4
Properties - GLP
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ GLP
Tab: Configuration
▶ Dosing units ▶ [Edit] ▶ Properties... ▶ Dosing unit - 'Na-
me'
▶ GLP
Tab: Configuration
▶ Solutions ▶ [Edit] ▶ Properties... ▶ Solution - 'Name' ▶ GLP
Last GLP test
Date of the last GLP test. This date can be selected by clicking on
in
the Select date dialog window (see Chapter 2.4.1, page 77).
Comment on GLP test
Comment regarding the GLP test.
Entry
1,000 characters
Monitoring of GLP validity
on | off (Default value: off)
If this check box is activated, then the time interval for the GLP test is
monitored.
Monitoring cannot be switched on until after a date has been entered in
the "GLP test date" field.
GLP test interval
Time interval to next GLP test. If a value is entered here, then the date in
the Next GLP test field will be adjusted automatically.
Only editable if monitoring is switched on.
Input range
1 - 999 days
Default value
999 days