6.6 Dosing units subwindow
■■■■■■■■■■■■■■■■■■■■■■
786
■■■■■■■■
viva 1.1 (for process analysis)
Dosing rate Dosing port 1
Rate at which dosing should take place via Dosing port 1. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see
below). When the function is carried out the rate will automatically be
decreased to the highest possible value.
Input range
0.01 - 166.00 mL/min
Default value
4 mL/min
Selection
maximum
Dosing rate Dosing port 2
Rate at which dosing should take place via Dosing port 2. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see
below). When the function is carried out, the rate will automatically be
decreased to the highest possible value.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Dosing rate Fill port
Rate at which dosing or filling should take place via Fill port. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used
(see below). When the function is carried out, the rate will automatically
be decreased to the highest possible value.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Dosing rate Special port
Rate at which dosing should take place via the Special port. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used
(see below). When the function is carried out, the rate will automatically
be decreased to the highest possible value.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Maximum dosing rate for the dosing unit in relation to the cylin-
der volume:
Cylinder volume
Maximum rate
2 mL
6.0 mL/min
5 mL
16.0 mL/min