Metrohm viva 1.1 (ProLab) User Manual
Page 958
7.11 Stepping Motor Controller
■■■■■■■■■■■■■■■■■■■■■■
946
■■■■■■■■
viva 1.1 (for process analysis)
Signal processing
Delayed edge
The input signal will only be recorded with a delay time on the rising or
falling edge.
Selection
ascending | descending
Default value
descending
Delay time
Time that has to elapse before the input signal on the rising or falling
edge is recorded.
Input range
0 - 214,783,647
Default value
0
Invert signal
on | off (Default value: off)
If this check box is activated, then the input signal received will be inver-
ted.
Message
Only active and editable for Type = Level detector.
Message by e-mail
on | off (Default value: off)
The message is sent to the address defined under [E-mail...] if this check
box is activated. The message is sent in text format.
[E-mail...]
[E-mail...] opens the Send e-mail window (see Chapter 2.5.1, page 79).
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message.
Regardless of which Action has been selected, the corresponding mes-
sage will also be displayed on the Messages tab of the live display (see
Chapter 3.8.4, page 161).
If the message is documented in the determination because of an Action,
it is displayed in the determination overview on the Messages tab in the
Information subwindow (see Chapter 4.7, page 356) and can be output
in a report.