Metrohm viva 1.1 (ProLab) User Manual
Page 556
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
544
■■■■■■■■
viva 1.1 (for process analysis)
Input range
0 - 999,999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula:
Input range
0 - 999,999 s
Default value
154 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.00 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 4,000 mAU
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed:
Max. waiting time
Maximum period of time for measurement.
Input range
0 - 999,999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.00 s