p = set pulse (pulse length = 1 s. If a pulse with a different length is to be
output, then a corresponding template must be defined for this purpose.)
Sample changers
Selection
Bit pattern with exactly 14 characters (0, 1, *,
p) | ************** | signal pattern
Default value
**************
894
Selection
Bit pattern with exactly 14 characters (0, 1, *,
p) | ************** | signal pattern
Default value
**************
Spectrometer
Selection
Bit pattern with exactly 10 characters (0, 1, *,
p) | ********** | signal pattern
Default value
**********
The output lines and bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Examples:
************1* sets the output line 1 to active (= set).
************0* sets the line to inactive.
NOTE
We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line statuses.
[Set]
Sets the bit pattern defined under Output signal.
8.4.5
Sampler changer functions
8.4.5.1
Sample changer - Overview
Program part: Manual control
If a tower of a sample changer is selected in the subwindow for the device
selection, then the functions and parameters that are possible with this
sample changer are displayed in the subwindow for functions/parameters.
Sample changer functions
The sample changer functions are displayed on the following tabs: