3 receive - wait for event/status – Metrohm viva 1.1 (ProLab) User Manual
Page 515
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
503
NOTE
A maximum of 10 event messages can be entered.
[Properties]
Opens the dialog window Wait for event/status to edit the event mes-
sage selected in the table which is to be waited for (see Chapter
5.5.2.6.5.3, page 503).
[Delete]
Deletes the selected event message.
Timeout
on | off (Default value: off)
If this option is activated, then a maximum waiting time for event mes-
sages can be entered. If none of the conditions defined above are fulfilled
after this time period, the command is canceled and the next command in
the sequence is executed.
Value
Input range
0 - 9,999.9
Default value
0
Unit
Selection
min | s
Default value
s
Message
This message defined here is displayed in the live window while waiting
for the result. The text editor for entering or editing the message is
opened with
or by double-clicking on the text field.
5.5.2.6.5.3
RECEIVE - Wait for event/status
Dialog window: Method
▶ RECEIVE ▶ Properties... ▶ [New]/[Properties] ▶ Wait for
event/status
Command
Name of the command whose event message or status is to be waited
for.
Selection
'Command name'