9 rls dos, 1 rls dos - overview, Rls dos – Metrohm viva 1.1 (ProLab) User Manual
Page 485
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
473
5.5.2.4.9
RLS DOS
5.5.2.4.9.1
RLS DOS - Overview
Dialog window: Method
▶ RLS DOS ▶ Properties... ▶ RLS DOS - 'Command name'
Command to release a dosing device for using it in other parallel run-
ning methods. It removes the reservation made by the current method.
Instruments
This command can be executed with the following instruments:
Voltammetry: 894
Dosing Interface: 846
Sample changer: 814, 815, 858, 919
Photometer: 089
Appearance
The command has the following appearance:
Parameters
The parameters for the RLS DOS command are configured in the follow-
ing dialog window:
■
Command variables
The following command variables are generated by the RLS DOS com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started