7.6 Avantes Spectrometer
■■■■■■■■■■■■■■■■■■■■■■
884
■■■■■■■■
viva 1.1 (for process analysis)
7.6.3
Properties - Settings
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Settings
Light source
Type
Type of light source used.
Selection
continuous | pulsed
Default value
continuous
Pulses per integration time
Frequency at which the lamp flashes. This parameter can be edited only if
Type = pulsed.
Input range
0 - 1,000
Default value
10
Analog output
Analog output
on | off (Default value: on)
If this check box is activated, then the measuring signal is output at the
analog output of the spectrometer. If the connected spectrometer is not
equipped with an analog output, then this setting will be ignored.
Sensitivity
Input range
1 - 214,783,647 mAU/mV
Default value
100 mAU/mV
7.6.4
Properties - Detector
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Detector
The Detector tab shows information on the detector that is built into the
spectrometer.
Detector type
Shows the detector type.
Minimum wavelength
Shows the minimum wavelength of the measurable range in nm.
Maximum wavelength
Shows the maximum wavelength of the measurable range in nm.