5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
528
■■■■■■■■
viva 1.1 (for process analysis)
Window = Time
Input range
0 - 999,999 s
Default value
0 s
Upper limit
Measured value for the upper limit of the window.
Time for the upper limit of the window.
Window = Measured value
Input range
0 - 4,000 mAU
Default value
4,000 mAU
Window = Time
Input range
0 - 999,999 s
Default value
999,999 s
5.5.2.7.4.5
MEAS Opt - Additional measured values
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ MEAS Opt - 'Command name' ▶ Addi-
tional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together with the measured values pres-
ent in the default settings in additional measured value columns.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this check box is activated, then a maximum of three new measured val-
ues can be calculated and saved from existing measured values that can
also be displayed as curves under the designation Calculated 1 - 3 and
be used as variables 'Command name.CA1 - 3' in formulas.
[New]
Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be entered (see Chapter
5.5.2.7.8.1, page 550).
[Properties]
Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be edited (see Chapter
5.5.2.7.8.1, page 550).