5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
488
■■■■■■■■
viva 1.1 (for process analysis)
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.5.2.5.4.3
CAL MEAS Opt - Measuring parameters
Tab: Method
▶ CAL MEAS Opt ▶ Properties... ▶ CAL MEAS Opt - 'Command
name'
▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The measuring parameters for the calibration are defined on this tab.
Parameters for the Avantes spectrometer
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
-1E+12 - 1E+12 (max. 10 digits) mAU
Default value
0 mAU
Integration time
Integration time for absorbance measurement.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are acquired and averaged.
Input range
1 - 10,000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.