5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
512
■■■■■■■■
viva 1.1 (for process analysis)
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
Avantes Spectrometer
Default value
Avantes Spectrometer
5.5.2.7.2.3
MEAS Ref - Measuring parameters
Tab: Method
▶ MEAS Ref ▶ Properties... ▶ MEAS Ref - 'Command name' ▶ Mea-
suring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Start wavelength
Lower limit of the spectrum.
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm
End wavelength
Upper limit of the spectrum.
Input range
100.0 - 2,000.0 nm
Default value
1,000.0 nm
Integration time
Integration time for the acquisition of the spectrum.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are acquired and averaged.