2 send - properties – Metrohm viva 1.1 (ProLab) User Manual
Page 511
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
499
Parameters
The parameters for the SEND command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the SEND command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
5.5.2.6.4.2
SEND - Properties
Dialog window: Method
▶ SEND ▶ Properties... ▶ SEND - 'Command name'
Command name
Name of the command.
Entry
25 characters
Send event messages
All of the defined event messages that are listed in the table are sent
simultaneously. The table cannot be edited directly.
NOTE
A maximum of 10 event messages can be entered.
[New]
Opens the Send event message dialog window for entering a new
event message (see Chapter 5.5.2.6.4.3, page 500).