Metrohm viva 1.1 (ProLab) User Manual
Page 331
■■■■■■■■■■■■■■■■■■■■■■
4 Database
viva 1.1 (for process analysis)
■■■■■■■■
319
For Calibration method = DT and Sample type = Standard, the fol-
lowing results will be displayed for each VA measuring command and
each substance:
'Command name','Substance name'.Calibration factor
Concentration of the substance in the sample.
'Command name','Substance name'.Calibration factorASD
Absolute standard deviation of the concentration of the substance in the
sample.
'Command name','Substance name'.Calibration factorRSD
Relative standard deviation of the concentration of the substance in the
sample.
Results for RC
For Calibration method = RC and Sample type = Sample, the follow-
ing results will be displayed for each VA measuring command and each
substance:
'Command name','Substance name'.Concentration
Concentration of the substance in the sample.
'Command name','Substance name'.Concentration.ASD
Absolute standard deviation of the concentration of the substance in the
sample.
'Command name','Substance name'.Concentration.RSD
Relative standard deviation of the concentration of the substance in the
sample.
The calibration data is stored in the database.
Results for LAT
For Calibration method = LAT and Sample type = Sample, the fol-
lowing results will be displayed for each VA measuring command and
each substance:
'Command name','Substance name'.Concentration
Concentration of the substance in the sample.
'Command name','Substance name'.Concentration.ASD
Absolute standard deviation of the concentration of the substance in the
sample.