2 meas spec - general/hardware – Metrohm viva 1.1 (ProLab) User Manual
Page 526
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
514
■■■■■■■■
viva 1.1 (for process analysis)
Identifica-
tion
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
DBL
Total duration for the processing of the command in s
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
PK{#}.ABS
Absorbance of the peak with the greatest absorbance
(maximum absorbance) for evaluation # (1 - 9)
PK{#}.CND
Intensity in dark spectrum 'Command
name'.PK{#}.WVL for evaluation # (1 - 9)
PK{#}.CNR
Intensity in reference spectrum 'Command
name'.PK{#}.WVL for evaluation # (1 - 9)
PK{#}.CNT
Intensity of the peak with the greatest absorbance
(maximum intensity) for evaluation # (1 - 9)
PK{#}.SAT
Specification as to whether the detector was saturated
with 'Command name'.PK{#}.WVL for evaluation # (1 -
9)
PK{#}.TRN
Transmission of the peak with the greatest absorbance
(minimum transmission) for evaluation # (1- 9) in %
PK{#}.WVL
Wavelength of the peak with the greatest absorbance
for evaluation # (1 - 9) of the discovered maximum
STY
Type of stop with which the command was stopped:
1 = finished normally, 0 = stopped manually
with [Quit] or [Stop]
5.5.2.7.3.2
MEAS Spec - General/Hardware
Tab: Method
▶ MEAS Spec ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument are defined on this tab.