5 Method
viva 1.1 (for process analysis)
■■■■■■■■
635
Selection
off
Default value
off
Waiting time 1
Waiting time during which the Potential 1 potential is applied to the
electrodes.
Only editable for Potential 1
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Potential 2
Potential that is applied to the electrodes during the Waiting time 2.
This potential can be used e.g. as deposition potential for the electro-
chemical deposition in stripping voltammetry.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Waiting time 2
Waiting time during which the Potential 2 potential is applied to the
electrodes.
Only editable for Potential 2
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Potential 3
Potential that is applied to the electrodes during the Waiting time 3.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Waiting time 3
Waiting time during which the Potential 3 potential is applied to the
electrodes.
Only editable for Potential 3
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s