Metrohm viva 1.1 (ProLab) User Manual
Page 886
7.5 814/815 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
874
■■■■■■■■
viva 1.1 (for process analysis)
ple positions (see Chapter 6.12.3, page 834) and special beakers (see
Chapter 6.12.3.2, page 834) are compared with the Min. beaker
radius. If this minimum beaker radius is not achieved, the run is stopped
and an error message is output. No check will be made if off is enabled.
Input range
1.0 - 100.0 mm
Selection
off
Default value
off
Lift rate
Lift rate for manual operation.
Input range
3 - 25 mm/s
Default value
25 mm/s
Axial distance
Distance between the axis of rotation of the sample rack and the swing
axis of the robotic arm.
814 USB Sample Processor, 858 Professional Sample Processor
Input range
100.0 - 300.0 mm
Default value
166.0 mm
815 Robotic USB Sample Processor XL
Input range
100.0 - 300.0 mm
Default value
196.0 mm
919 IC Autosampler plus
Input range
100.0 - 300.0 mm
Default value
189.0 mm
Swing Head
This parameter is only shown only if a Swing Head is mounted on the
tower.
Serial number
Shows the serial number of the Swing Head.
Swing position
Lift position that is valid for all four external positions and at which the
robotic arm turns to the external positions.
Input range
0 - 235 mm
Default value
0 mm