5 Method
viva 1.1 (for process analysis)
■■■■■■■■
579
Current view
If this option is enabled, then the view selected on the workplace
remains open when the determination is started.
View
If this option is enabled, the selected workplace view is automatically
opened when the determination is started.
If the View option is selected, then an existing workplace view can be
chosen from the selection list.
Selection
'Workplace view'
NOTE
The selected workplace view is only opened when a single determina-
tion or the first determination of a series is started, but not when a new
method is loaded within a determination series.
Track view for live window
Live display 1
Selection of the track to be displayed in the Live display 1 subwindow.
Selection
Main track | Selection from existing tracks
Default value
Main track
Live display 2
Selection of the track to be displayed in the Live display 1 subwindow.
Selection
Main track | Selection from existing tracks
Default value
Main track
Electrode test for VA measuring commands
Electrode check
on | off (Default value: on)
If this check box is activated, then an electrode test will be performed for
the reference and auxiliary electrodes before the first VA measuring com-
mand is carried out for methods with VA measuring commands. This test
checks whether an electrode is connected and ready for operation. An
electrode test is performed only once for each determination.