5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
534
■■■■■■■■
viva 1.1 (for process analysis)
5.5.2.7.5.3
MEAS Opt Conc - Measuring parameters
Tab: Method
▶ MEAS Opt Conc ▶ Properties... ▶ MEAS Opt Conc - 'Command
name'
▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Tab for Avantes spectrometer
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded.
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
–1E+12 - 1E+12 (max. 10 digits)
Default value
0.0
Entry
1 - 10 characters
Selection
mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L
Default value
ppm
Integration time
Integration time for absorbance measurement.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
Number of spectra that are recorded and averaged.
Input range
1 - 10,000
Default value
10
Smoothing
Number of neighboring pixels with which the light intensity of each mea-
surement pixel is averaged.