Metrohm viva 1.1 (ProLab) User Manual
Page 957
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
viva 1.1 (for process analysis)
■■■■■■■■
945
The port type of the first digital input is always of the Emergency stop
type and cannot be modified.
IO Controller
Selection
General input | Emergency stop | Level detec-
tor | Leak detector
Default value
General input
Stepping Motor Controller
Selection
General input | Emergency stop
Default value
General input
NOTE
Each stepping motor has a port of the Type:
■
Reference position
■
Safety switch 1
■
Safety switch 2
■
Safety switch 3
The port information for these port types is automatically read out of
the Stepping Motor Controller.
Assigned ports
Selection of an output port assigned to an input port. The signal of the
output port depends on the signal of the input port.
Only active and editable for the port types Level detector and Leak
detector.
Selection
All digital outputs
NOTE
Digital outputs of the Type = Alarm signal cannot be selected.
Signal
Selection of the value which the signal on the output port should have.
Only active and editable for the port types Level detector and Leak
detector.
Selection
Input signal | Invertet input signal
Default value
Input signal