5.6 Evaluation subwindow
■■■■■■■■■■■■■■■■■■■■■■
662
■■■■■■■■
viva 1.1 (for process analysis)
Manual
If this option is selected, then the base points can be specified man-
ually for the baseline calculation.
CPVS
Selection
Automatically | Manual
Default value
Manual
Automatically
If this option is selected, then the base points are determined automat-
ically for the baseline calculation.
Manual
If this option is selected, then the base points can be specified man-
ually for the baseline calculation.
Start base point
Manually specified start base point at which the baseline is to begin.
CVS
Input range
–5.000 - 5.000 V
Default value
0.0 V
CPVS
Input range
0 - t s
Default value
3.9 s
t is the total duration of data recording for curve
evaluation. It is the sum of plating time + stripping
time 1 + stripping time 2 + stripping time 3.
End base point
Manually specified end base point at which the baseline is to end.
CVS
Input range
–5.000 - 5.000 V
Default value
0.0 V
CPVS
Input range
0 - t s
Default value
6.5 s
t is the total duration of data recording for curve
evaluation. It is the sum of plating time + stripping
time 1 + stripping time 2 + stripping time 3.