5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
646
■■■■■■■■
viva 1.1 (for process analysis)
5.5.2.11.4.3
CP - Pretreatment
Tab: Method
▶ CP ▶ Properties... ▶ Pretreatment
Command name
Name of the command.
Entry
25 characters
The parameters for the pretreatment can be set on this tab.
Stirring time
Waiting time during which the solution is stirred in the measuring cell.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Potential 1
Potential that is applied to the electrodes during the Waiting time 1.
This potential can be used e.g. as cleaning potential for the electrochemi-
cal cleaning of solid-state electrode surfaces that have been contaminated
with reaction products of electrode redox processes.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Waiting time 1
Waiting time during which the Potential 1 potential is applied to the
electrodes.
Only editable for Potential 1
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Potential 2
Potential that is applied to the electrodes during the Waiting time 2.
This potential can be used e.g. as deposition potential for the electro-
chemical deposition in stripping voltammetry.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Waiting time 2
Waiting time during which the Potential 2 potential is applied to the
electrodes.
Only editable for Potential 2
≠ off, otherwise inactive.