5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
484
■■■■■■■■
viva 1.1 (for process analysis)
Selection
Linear regression | Quadratic regression |
cubic | automatic
Default value
Linear regression
Confidence interval
Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers; they
are not integrated in the calculation of the calibration coefficient.
Selection
99 | 98 | 95 | off
Default value
98
off
The confidence interval will not be checked.
5.5.2.5.4
CAL MEAS Opt
5.5.2.5.4.1
CAL MEAS Opt - Overview
Dialog window: Method
▶ CAL MEAS Opt ▶ Properties... ▶ CAL MEAS Opt - 'Com-
mand name'
Command for measuring standard solutions for the calibration of
colorimetric sensors, used together with the CAL LOOP Opt com-
mand.
Instruments
This command can be executed with the following instruments:
Spectrometer: Avantes
Photometer: 089
Appearance
The command has the following appearance:
Parameters
The parameters for the CAL MEAS Opt command are set on the follow-
ing tabs:
■
General/Hardware
Selection of instrument type and colorimetric sensor.
■
Measuring parameters
Parameters for setting the measurement procedure.