5 solutions subwindow, 1 solutions - general, 2 solution table – Metrohm viva 1.1 (ProLab) User Manual
Page 786: 1 solution table, Solutions subwindow, Solutions - general, Solution table, Solutions
6.5 Solutions subwindow
■■■■■■■■■■■■■■■■■■■■■■
774
■■■■■■■■
viva 1.1 (for process analysis)
Device name
Assignment of the required device to the command. The list of all devices
of the required device type present in the configuration is available in the
combo box.
6.5
Solutions subwindow
6.5.1
Solutions - General
Subwindow: Configuration
▶ Solutions
Solutions subwindow
The Solutions subwindow contains the solution table with all auxiliary
and standard solutions. It can be shown in the Configuration program
part as a part of the configuration view or (if not present on the user inter-
face) in a separate window with View
▶ Quick access. The subwindow
can be enlarged and reduced as required; it can also be maximized.
Standard solutions
Standard solutions are used as solutions for the dosing commands ADD
STD, ADD AUX, LQH or PREP. For the addition with the ADD STD
command, they have to be additionally defined as standards in the
method already.
Auxiliary solutions
Auxiliary solutions are used as solutions for the dosing commands ADD
AUX, LQH or PREP.
6.5.2
Solution table
6.5.2.1
Solution table
Subwindow: Configuration
▶ Solutions
Contents
The following information about the auxiliary and standard solutions is
displayed in the solution table by default:
Solution name
Name of the solution.
Solution type
Type of the solution.
Manufacturer
Name of the manufacturer of the solution.