6.11 Global variables subwindow
■■■■■■■■■■■■■■■■■■■■■■
830
■■■■■■■■
viva 1.1 (for process analysis)
[Limit values]
Opens the Limit values for Global Variable dialog window, in which
the warning and intervention limits can be defined for the global variable.
These limits apply only to the graph; no monitoring takes place during
value assignment.
[Delete history]
Deletes the history.
6.11.3.4
History - Limit values
Dialog window: Configuration
▶ Global Variables ▶ [Edit] ▶ Properties... ▶ Global
variable - 'Name'
▶ History ▶ [Limit values] ▶ Limit values global variable -
'Name'
In the Limit values for Global Variable dialog window both warning
and intervention limits can be defined for the value of the global variable.
If you have defined the limits, then these will be shown in the graph in
orange for warning limits and red for intervention limits. However,
whether these limits are observed is not monitored, i.e., a breach of these
limits does not trigger any action.
Warning limits for Global Variable
on | off (Default value: off)
If this check box is activated, then the values of the global variable will be
shown in orange if these limits are breached.
Lower limit
Lower warning limit for global variable.
Input range
-1.0E+99 - 1.0E+99 (max. 10 digits)
Upper limit
Upper warning limit for global variable.
Input range
-1.0E+99 - 1.0E+99 (max. 10 digits)
Intervention limits for Global Variable
on | off (Default value: off)
If this check box is activated, then the values of the global variable will be
shown in red if these limits are breached.
Lower limit
Lower warning limit for global variable.
Input range
-1.0E+99 - 1.0E+99 (max. 10 digits)