Metrohm viva 1.1 (ProLab) User Manual
Page 645
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
633
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
894 Professional CVS
Default value
894 Professional CVS
Sensors/Electrodes
Working electrode
Selection of an electrode of the Working electrode type from the elec-
trodes of the RDE type present in the electrode table.
Selection
RDE | 'Electrode name'
Default value
RDE
RDE
Standard electrode of the RDE type, which cannot be deleted.
'Electrode name'
Electrode of the RDE type from the electrode table.
Sensor type
Shows the type of working electrode (RDE).
Reference electrode
Selection of an electrode of the Reference electrode type from the elec-
trodes available in the electrode table.
Selection
Reference electrode | 'Electrode name'
Default value
Reference electrode
Reference electrode
Standard electrode of the Reference electrode type, which cannot
be deleted.
'Electrode name'
Electrode of the Reference electrode type from the electrode table.
Auxiliary electrode
Selection of an electrode of the Auxiliary electrode type from the elec-
trodes available in the electrode table.
Selection
Auxiliary electrode | 'Electrode name'
Default value
Auxiliary electrode