5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
540
■■■■■■■■
viva 1.1 (for process analysis)
■
General/Hardware
Parameters for instrument.
■
Measuring parameters
Parameters for setting the measurement procedure.
Command variables
The following command variables are generated by the MEAS TMF com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identifica-
tion
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has never
been started
DBL
Total duration for the processing of the command in s
EME
End measured value (measured value after processing of
the command) in the unit of the measured value
ETE
End temperature (temperature after the command has
been processed) in °C
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
invalid (variable not available) = Command has never
been started
IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
ITE
Initial temperature (temperature before start conditions
are processed) in °C
LP.MEA
Measured value for the last measuring point in the mea-
suring point list in the unit of the measured value
LP.TEM
Temperature for the last measuring point in the measur-
ing point list in °C
LP.TIM
Time in s until the last measuring point in the measuring
point list is reached