5 Method
viva 1.1 (for process analysis)
■■■■■■■■
621
5.5.2.11.2.5
CVS - Post-treatment
Tab: Method
▶ CVS ▶ Properties... ▶ Post-treatment
Command name
Name of the command.
Entry
25 characters
Cleaning
Cleaning potential
Potential that is applied to the electrodes during the Cleaning time. This
potential can be used e.g. for the electrochemical cleaning of solid-state
electrode surfaces that have been contaminated with reaction products of
electrode redox processes.
Input range
–5.00 - 5.00 V (Increment: 0.001)
Selection
off
Default value
off
Cleaning time
Waiting time during which the Cleaning potential is applied to the elec-
trodes.
Only editable for Cleaning potential
≠ off, otherwise inactive.
Input range
0.0 - 9999.9 s (Increment: 0.1)
Default value
0.0 s
Standby potential
Standby potential
The standby potential is applied to the electrodes after completion of the
sweep. It remains in effect until it is switched off manually or until a new
potential is applied to the electrodes. If no standby potential is defined,
then the potential applied to the electrodes is automatically switched off
after completion of the sweep.
Input range
–5.00 - 5.00 V (Increment: 0.001)
Selection
off
Default value
off