5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
630
■■■■■■■■
viva 1.1 (for process analysis)
Result variables if called by CALL ELECTROLYTE
The following result variables are generated in the method run when a
CPVS command is called by CALL ELECTROLYTE and can be used in
formulas:
Result variables per substance, electrolyte and replication:
Designation 'RS.'Command name'.'Substance name'.Electro-
lyte.REP{y}.Variable identification'
Identification
Description
SUBST
Substance name
POT
Stripping potential
AREA
Peak area
HGT
Peak height
START
Start base point, where the baseline of the peak eval-
uation begins
END
End base point, where the baseline of the peak eval-
uation ends
AREANORM
Standardized peak area (AREA/Electro-
lyte.AREA.MNV)
HGTNORM
Standardized peak height (HGT/Electro-
lyte.HGT.MNV)
Result variables per substance and electrolyte
Designation 'RS.'Command name'.'Substance name'.Electro-
lyte.Variable identification'
Identification
Description
AREA.MNV
Mean value of the peak areas of all replications
AREA.ASD
Absolute standard deviation of the peak areas of all
replications
AREA.RSD
Relative standard deviation of the peak areas of all
replications
HGT.MNV
Mean value of the peak heights of all replications
HGT.ASD
Absolute standard deviation of the peak heights of
all replications
HGT.RSD
Relative standard deviation of the peak heights of all
replications