3 cvs - pretreatment, Pretreatment – Metrohm viva 1.1 (ProLab) User Manual
Page 629
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
617
Stirrer
Stirring rate
Number of revolutions per minute of the stirrer.
Input range
0 - 3000 min
⁻¹ (Increment: 100)
Default value
2500 min
⁻¹
Hydrodynamic measurement
on | off (Default value: on)
If this check box is activated, then stirring will continue during the sweep.
Only editable for Sensor type = RDE, otherwise disabled.
5.5.2.11.2.3
CVS - Pretreatment
Tab: Method
▶ CVS ▶ Properties... ▶ Pretreatment
Command name
Name of the command.
Entry
25 characters
The parameters for the pretreatment can be set on this tab.
Stirring time
Waiting time during which the solution is stirred in the measuring cell.
Input range
0.0 - 9999.9 s (Increment: 0.1)
Default value
5.0 s
Potential 1
Potential that is applied to the electrodes during the Waiting time 1.
This potential can be used e.g. as cleaning potential for the electrochemi-
cal cleaning of solid-state electrode surfaces that have been contaminated
with reaction products of electrode redox processes.
Input range
–5.00 - 5.00 V (Increment: 0.001)
Selection
off
Default value
off
Waiting time 1
Waiting time during which the Potential 1 potential is applied to the
electrodes.
Only editable for Potential 1
≠ off, otherwise inactive.
Input range
0.0 - 9999.9 s (Increment: 0.1)
Default value
0.0 s