Metrohm viva 1.1 (ProLab) User Manual
Page 596
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
584
■■■■■■■■
viva 1.1 (for process analysis)
text field (see Chapter 2.4.2, page 78). The formula editor can also be
opened within the text field (see Chapter 2.3, page 21).
Entry
Text (unlimited)
Display message
on | off (Default value: on)
If this check box is activated, then all active tracks will be stopped and the
message defined in the text field will be displayed if the limit value is
breached.
Record message
on | off (Default value: on)
If this check box is activated, then the message defined in the text field
will be documented in the determination.
Message by e-mail
on | off (Default value: off)
If this check box is activated, then the message defined in the text field
will be output to the address defined under [E-mail...] if the limit value is
breached.
[E-mail...]
Opens the Send e-mail window for defining the e-mail parameters (see
Chapter 2.5.1, page 79).
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message if the limit value is breached.
Action
NOTE
Is only displayed for sample data variables of the Number or Date/
Time type.
Action
on | off (Default value: off)
If this option is enabled, then one of the two following actions will auto-
matically be executed if the limit values are breached.