5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
636
■■■■■■■■
viva 1.1 (for process analysis)
5.5.2.11.3.4
CPVS - Sweep
Tab: Method
▶ CPVS ▶ Properties... ▶ Sweep
Command name
Name of the command.
Entry
25 characters
Equilibration potential
Potential applied during the Equilibration time.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Default value
0.450 V
Equilibration time
Time during which the Equilibration potential is applied to the elec-
trodes.
Input range
0.0 - 50.0 s (Increment: 0.1)
Default value
5.0 s
Plating potential
Potential applied during the Plating time.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Default value
–0.175 V
Plating time
Time during which the Plating potential is applied to the electrodes.
Input range
0.0 - 50.0 s (Increment: 0.1)
Default value
4.0 s
Stripping
In this table, you can define up to five stripping potentials, which are suc-
cessively applied during the corresponding stripping time. By default, three
stripping potentials are specified.
Stripping potential
Potential applied during the Stripping time.
Stripping time
Time during which the Stripping potential is applied to the electrodes.