Metrohm viva 1.1 (ProLab) User Manual
Page 46
2.3 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
viva 1.1 (for process analysis)
Identifica-
tion
Description
Commands
LP.EXx
External value x (1 - 3) for the last measuring point on the measuring point list
MEAS Opt,
MEAS Opt
Conc
LP.MEA
Measured value for the last measuring point in the measuring point list in the unit
of the measured value
CAL MEAS
Opt, MEAS
Opt, MEAS
Opt Conc,
MEAS TMF
LP.TEM
Temperature for the last measuring point in the measuring point list in °C
CAL MEAS
Opt, MEAS
Opt, MEAS
Opt Conc,
MEAS TMF
LP.TIM
Time in s until the last measuring point in the measuring point list is reached
CAL MEAS
Opt, MEAS
Opt, MEAS
Opt Conc,
MEAS TMF
LP.TRN
Transmission of the last measuring point in the measuring point list in %
CAL MEAS
Opt, MEAS
Opt, MEAS
Opt Conc
LPO
Current absolute lift position in mm (entry when ending the command)
LIFT
LST
Start time of the loop command (Date/Time)
CALL COND
CAL LOOP
Opt, LOOP
MA.CNT
Intensity for maximum measured value in counts
MEAS Opt
MA.MEA
Maximum measured value in the unit of the measured value
MEAS Opt
MA.TEM
Temperature for the maximum measured value in °C
MEAS Opt
MA.TIM
Time in s until the maximum measured value is reached
MEAS Opt
MA.TRN
Transmission for maximum measured value in %
MEAS Opt
MI.CNT
Intensity for minimum measured value in counts
MEAS Opt
MI.MEA
Minimum measured value in the unit of the measured value
MEAS Opt
MI.TEM
Temperature at minimum measured value in °C
MEAS Opt
MI.TIM
Time for the minimum measured value in s
MEAS Opt
MI.TRN
Transmission for minimum measured value in %
MEAS Opt