3 send - sending the event message – Metrohm viva 1.1 (ProLab) User Manual
Page 512
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
500
■■■■■■■■
viva 1.1 (for process analysis)
[Properties]
Opens the Send event message dialog window for editing the event
message selected in the table(see Chapter 5.5.2.6.4.3, page 500).
[Delete]
Deletes the selected event message.
5.5.2.6.4.3
SEND - Sending the event message
Dialog window: Method
▶ SEND ▶ Properties... ▶ [New]/ [Properties] ▶ Send event
message
Receiver
Selection of the receiver type.
Selection
System | Command
Default value
System
Selection
Selection of the recipient address. All of the command names already
defined in the method are available when Receiver = Command. This
field cannot be edited when Receiver = System.
Selection
'Command name' | 'first command'
Default value
'first command'
Event message
Selection of the event message that is to be sent. The following event
messages can be selected, depending on the recipient:
System
Selection
Quit | Hold | Stop
Default value
Quit
Command
Selection
Quit | Hold | Continue
Default value
Quit
Receiver
Selection
Event
message
Meaning
System
-
Hold
Stop determination (all tracks). This corresponds
to the [Hold] button in the Run dialog win-
dow in the Workplace program part.
System
-
Quit
Stop determination (series continues running).