2 set total volume - properties, 5 calibration commands, 1 calibration command - overview – Metrohm viva 1.1 (ProLab) User Manual
Page 488
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
476
■■■■■■■■
viva 1.1 (for process analysis)
Command variables
The following command variables are generated by the command SET
TOTAL VOLUME in the method run and can be used in formulas under
the designation 'Command name.Variable identification':
Identification
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has
never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
invalid (variable not available) = Command has
never been started
5.5.2.4.10.2
SET TOTAL VOLUME - Properties
Dialog window: Method
▶ SET TOTAL VOLUME ▶ Properties... ▶ SET TOTAL VOL-
UME - 'Command name'
Command name
Name of the command.
Entry
25 characters
Total volume
Volume
Quantity of the current total volume in the measuring vessel, e.g. after
automatic rinsing or after a manual exchange of solution.
Input range
0.000 - 1,000.000 mL
Default value
0.000 mL
5.5.2.5
Calibration commands
5.5.2.5.1
Calibration command - Overview
Menu item: Method
▶ Insert ▶ New command... ▶ Calibration
The following calibration commands can be selected: