5 cpvs - post-treatment, 6 cpvs - potentiostat, Post-treatment – Metrohm viva 1.1 (ProLab) User Manual
Page 650: Potentiostat
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
638
■■■■■■■■
viva 1.1 (for process analysis)
Calculation: (Preparation cycles + Measuring cycles)
× (Equilibration
time + Plating time + Stripping time 1 - 5 + Cleaning time)
NOTE
The actual current measurement takes place during the last quarter of
the measuring interval.
5.5.2.11.3.5
CPVS - Post-treatment
Tab: Method
▶ CPVS ▶ Properties... ▶ Post-treatment
Command name
Name of the command.
Entry
25 characters
Standby potential
Standby potential
The standby potential is applied to the electrodes after completion of the
sweep. It remains in effect until it is switched off manually or until a new
potential is applied to the electrodes. If no standby potential is defined,
then the potential applied to the electrodes is automatically switched off
after completion of the sweep.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
5.5.2.11.3.6
CPVS - Potentiostat
Tab: Method
▶ CPVS ▶ Properties... ▶ Potentiostat
Command name
Name of the command.
Entry
25 characters
Current measuring range
Maximum range
Limitation of the highest current measuring range.
Selection
200 pA | 2 nA | 20 nA | 200 nA | 2
µA | 20 µA |
200
µA | 2 mA | 20 mA | 224 mA
Default value
224 mA