Metrohm viva 1.1 (ProLab) User Manual
Page 639
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
627
Command variables
The following command variables are generated by the command CPVS
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has
never been started
DBL
Total duration for the processing of the command in
s
ELT
Variable for electrode test:
0 = Electrode test has not been performed
1 = Electrode test has been performed once
2 = Electrode test has been performed twice
3 = Electrode test has been performed three times
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
invalid (variable not available) = Command has
never been started
HR
Highest current measuring range used:
1 = 224 mA | 3 = 20 mA | 5 = 2 mA | 7 = 200 µA |
9 = 20 µA | 11 = 2 µA | 13 = 200 nA | 15 = 20 nA |
17 = 2 nA | 19= 200 pA
LR
Lowest current measuring range used:
19= 200 pA | 17 = 2 nA | 15 = 20 nA | 13 = 200 nA
| 11 = 2 µA | 9 = 20 µA | 7 = 200 µA | 5 = 2 mA |
3 = 20 mA | 1 = 224 mA
NMP
Number of measuring points in measuring point list
RPM
Stirring rate in min
–1
STY
Type of stop with which the command was stopped:
1 = finished normally, 0 = stopped manually
with [Quit] or [Stop]